How to evaluate an Out-of-control result of Internal Quality Control before repeating it?

Dr. Mehmood ul Hasan Malik


Case Vignette:
A lab technician informed the haematologist about a violation in Internal Quality Control (IQC) result. The haematologist asked the technician to bring the print outs of the results. The result showed letter L before the results of WBC and Neutrophils; no flag was seen before RBC, Hb or Platelets.
The technician was eager to repeat it. But the haematologist stopped her and asked her to follow the correct line of action before repeating the IQC. Upon asking the details, he advised her to investigate the issues in the IQC vials first and then any other ones. Regarding IQC vials, the volume was adequate. There was no clot and the technician was sure about proper mixing before running IQC. She also gave around 15 minutes for the vials to come to the room temperature before running the test. The vials were not expired. They were always stored in the refrigerator at 4 degree centigrade and the refrigerator log showed no malfunction. The external thermometre was also showing the similar temperature as was shown on the screen. The technician didn't write the date of opening the vials. As the vials were ready-to-use type, so there was no possibility of mistake in reconstitution.
Suspecting any stability issue, the haematologist asked the technician to open new vials and to write down the date of opening the vials and asked her to run IQC again to see if stability was the issue; the result was the same! The haematologist guided her that this repetition is not the repetition which she was trying to do as a first step as now the investigation was going on instead of only trying to get a good result by luck!
No change in result excluded any instability issue. Now, the haematologist asked her to see any abnormality in reagents, routine maintenance, Planned Preventive Maintenance (PPM), and External Quality Assurance (EQA). The technician had done daily maintenance, and the PPM was also not pending. The EQA has passed too.
The haematologist suspected any issue in Fluidics or Flowcell or Sample Rotatory Valve or Aperture or Aperture Clogging of WBC channel and asked her to perform cleaning of WBC channel. She performed all of them but repeat testing did not show any improvements after each step.
Calibrator


We can classify the out-of-control results into following:
1. UNACCEPTABLE RESULT IN ONE LEVEL:
   i. Affecting one analyte
   ii. Affecting more than one analyte
2. UNACCEPTABLE RESULT IN MORE THAN ONE LEVEL:
   i. Affecting one analyte
   ii. Affecting more than one analyte